The fully optioned Phosentix MicroQC™ Plate has every option available on the plate – fluorescence, absorbance (NIST linked), luminescence, kinetic luminescence simulations, crosstalk performance, and incubator temperature monitoring. The full specifications are shown below.
Analytical Specifications:
Twelve absorbance standards ranging from 0.045 OD to 2.50 OD, NIST traceable
Holmium Oxide Glass Filter for wavelength accuracy testing
Twelve precision diffused luminescence standards
Twelve solid-state fluorescence standards
Integrated temperature sensor with 0.2C accuracy for monitoring of microplate reader incubation
Assay simulation luminescence well with ramped luminescence value to simulate time varying luminescence.
Crosstalk luminescence well, blank well surrounded by four luminescence standards, allows performance metrics for 96-well crosstalk.
1536 corner alignment wells based on SBS 1536 microplate standards
Temperature measurement, NIST traceable
X-Y attitude measurement with accelerometers for proper optical alignment
Chassis Specifications:
ePaper display indicating operating mode, temperature, calibration date, calibration due date, serial number, instrument serial number, and battery level
USB port for battery charging
Li-ion battery with battery charge indicator
Lightweight aluminum chassis, precision machined and black anodized for minimum light reflection.
Entire assembly machined on precise SBS standard dimensions to 0.1mm tolerances
The Phosentix MicroQC Plate will work with readers from Berthold™, Biotek™, BMG LABTECH™, Molecular Devices™, Perkin Elmer™, Tecan™, Thermo™, and many others.